C (E3) 08/2015 page 65/93
Recommended procedure:
Turn off the unit
Pull out the CO
2
sensor without any rotation from the connection socket in the upper part of the rear
and remove it from the inner chamber.
The CO
2
sensor is temperature resistant up to a maximum temperature of 60 °C / 140 °F. It is
therefore required to remove it before performing a hot-air sterilization.
CAUTION
Excess temperature.
Damage to the CO
2
sensor.
Do NOT expose the CO
2
sensor to hot-air sterilization.
The CO
2
sensor head was especially adjusted for your specific chamber. To avoid confusion, an ad-
hesive label with a serial number is adhered to the sensor head. When exchanging the sensor, repeat
CO
2
control adjustment.
CAUTION
Different CO
2
sensor.
Invalid calibration.
Do NOT change the CO
2
sensor head.
Write to down the serial number of the CO
2
sensor.
Empty the water pan.
WARNING
Danger of implosion.
Damage to the unit.
Empty the water pan before starting the hot-air sterilization.
Clean the C 170.
Water pan, shelf holder, and shelves must be inside the incubator, the water pan at its usual place on
the bottom.
Close the inner glass door and the outer unit door.
Turn on the unit.
Activate the sterilization procedure: Press the controller button “180 °C” for 3 seconds to access the
sterilization program.