C (E3) 08/2015 page 80/93
19.4 C 170 technical data
Exterior dimensions
Width
mm / inch
740 / 29.13
Height (incl. feet)
mm / inch
870 / 34.25
Depth
mm / inch
715 / 28.15
Depth including door handle, I-triangle, and
connections
mm / inch 795 / 31.30
Wall clearance rear
mm / inch
100 / 3.94
Wall clearance side
mm / inch
50 / 1.97
Number of doors
1
Number of inner glass doors
1
Interior dimensions
Width
mm / inch
560 / 22.05
Height
mm / inch
600 / 23.62
Depth
mm / inch
500 / 19.69
Interior volume
l / cu.ft.
168 / 5.93
Number of shelves, series / max.
3/8
Size of shelf, width x depth
mm x mm
inch x inch
538 x 447
21.18 x 17.60
Temperature data
Temperature range, 7 °C / 12.6 °F above ambient,
up to
°C / °F 50 / 122
Temperature fluctuation
± K
0.1
Temperature uniformity (variation) at 37 °C/ 98.6 °F
± K
0.3
Recovery time after door open for 30 sec
at 37 °C / 98.6°F
minutes 6
Humidity data
Humidity range
% r.H.
90 to 95
CO
2
data
CO
2
range
vol.-% CO
2
0 to 20
Setting accuracy
vol.-% CO
2
0.1
Recovery time after door open for 30 sec
at 5 vol.-% CO
2
minutes 5
CO
2
measurement
Drift-free CO
2
infrared absorption measurement
system
Connection hose nozzle DN6 for CO
2
Connection to the unit for hose with internal diameter
mm / inch 6 / 0.24
Electrical data
(model version C170-230V-R)
IP system of protection acc. to EN 60529
IP
20
Nominal voltage
(+/-10%)
at 50 Hz power frequency
V
230
at 60 Hz power frequency
V
230
Current type
1N~
Nominal power
kW
1.10
Energy consumption at 37 °C / 98.6°F
Wh/h
90
Power plug (IEC connector plug)
Shock-proof plug
Installation category acc. to IEC 61010-1
II
Pollution degree acc. to IEC 61010-1
2
Unit fuse
5x20mm / semi time-lag / 10 A