N-13Rev F
InStallatIon
N-13
characterS aux. Id
This setting will set the minimum and maximum number of characters in the Auxiliary ID.
If the setting of aux. Id enaBle/dISaBle is enaBled, an Auxiliary ID is mandatory for each
patient test.
1. From a list of programmable parameters, choose meter SettIngS
using the
keys.
2. Press the key.
3. Use the
keys to select
Id SettIngS.
4. Press the key.
5. Use the
keys to select the number to the right of characterS aux. Id under
the mIn column.
6.
Use the
keys to
choose the smallest number of characters that will
appear in the Auxiliary ID. The minimum number of characters possible is 1.
7. Use the key to select the number to the right of characterS aux. Id under the
max column.
8.
Use the
keys to
choose the largest number of characters that will appear
in the Auxiliary ID. The maximum number of characters possible is 12.
9. Press the key to save changes.
aux. Id enaBle/dISaBle
This setting will turn on or turn off the Auxiliary ID function. The Auxiliary ID may be used
for recording a test order number, physician ID or other specialized identifying information.
1. From a list of programmable parameters, choose meter SettIngS
using the
keys.
2. Press the key.
3. Use the
keys to select
Id SettIngS.
4. Press the key.
5. Use the
keys to select the word to the right of
aux. Id enaBle/dISaBle.
6.
Use the
keys to
choose dISaBled or enaBled. If dISaBled is selected, the
Auxiliary ID prompt screens will never appear. If enaBled is selected, an Auxiliary ID
is mandatory for each patient test.
7. Press the key to save changes.
maIn menu
Set parameterS
meter SettIngS
Id SettIngS
Note: The
AuxiliaryIDis
shown only on the
NewResultdisplayor
printout. It is not
visible in recalled
results. Tracking
may be managed
through the optional
software package,
Triage Census
Data Management.