M-6Rev F
care & maIntenance
Service and Maintenance Procedures
Service and Maintenance Procedures
If any service or maintenance is required, the Triage MeterPro Meter should be sent to
the manufacturer. No maintenance other than paper / battery replacement and periodic
external cleaning is required of the operator.
Paper Replacement
Changing Paper Between Functions
1. Tear off any excess paper sticking out of the Triage MeterPro Meter.
2. Remove the paper compartment cover by pulling up on the cover as indicate by
the arrow on the back of the cover.
3. Remove unused paper or the empty paper spindle from paper compartment.
4. Tear or cut a clean, straight edge to feed into the printer. Do not cut paper at an
angle, as the printer must sense the edge of the paper along the feed path.
5. Insert the new roll of paper into the paper compartment.
6. Position the paper such that the paper will feed from under the roll (as opposed
to over the top of the roll, see picture, above).
7. Insert the paper edge under the paper roller (platen) until it firmly seats or
resistance is felt.
8. Press the key.
9. Replace the cover of the printer and continue operation.
Note: The
printer contains
a paper sensor and
will feed the new
paper roll only when
paper with a clean
straight line is pressed
into the paper roller.
Note: The printer
utilizes thermal
paper, if the printed
paper is blank after
replacing a roll, verify
that the paper roll is
not upside down.
Note: If PAPER
FEED button is
pressed before the
printer senses the
paper, the printer will
not respond to the
command.