N-25Rev F
InStallatIon
N-25
rangeS
Depending on the panel type, the range on the display and printout will be titled
reFerence rangeS or threSholdS. The low end, when the test displays thresholds,
is set to ‘0’ and is unchangeable.
Some analytes are designed for three distinct ranges. The meter distinguishes between
two range and three range analytes in the setup screen by differentiating the symbol
between the lower and upper values.
• Analytes with a dash Two ranges. The normal range is any value between
the two numbers displayed. Depending on the product, the normal range may be
inclusive or exclusive of the upper number. Consult the product package insert or
call Biosite Technical Services for clarification. These values will be in reverse video
on the patient results.
• Analytes with a comma – Three ranges possible. If the lower number is 0.0, the
meter treats the analyte as having only two ranges. If the lower number is not 0.0,
the normal range is any value less than the lower number. The first abnormal range
is any value between the two numbers and is inclusive of these numbers. These
values will be boxed on patient results. The second abnormal range is any value
greater than the upper number. These values will be in reverse video on patient
results. Consult the product package insert or call Biosite Technical Services for
clarification.
1. From the list of programmable parameters, choose rangeS
using the
keys.
2. Press the key.
3. Select the test panel type using the
keys.
4. Select the analyte and level using the
keys. The hIgh and low value
for each of the analytes can be changed using the keypad to type in the
values. To correct, press the key to clear the entire number. Then type in the
correct number.
5. Press the key to save changes.
maIn menu
Set parameterS
Note: The
field will be
disabled if the panel
settingisfixed.
Consult the specific
package insert to
determine if the
setting may be
changed.