N-29Rev F
InStallatIon
N-29
5. Press the key to save changes.
numBer oF controlS
1. From the list of programmable parameters, choose Qc parameterS
using the
keys.
2. Press the key.
3. Use the
keys to select the information to the right of numBer oF controlS.
4. Using the
keys, select one of these options: 1 or 2.
5. Press the key to save changes.
numBer oF control Standard devIatIonS For QuantItatIve teStS
1. From the list of programmable parameters, choose Qc parameterS
using the
keys.
2. Press the key.
3. Use the
keys to select the information to the right of num Quant Std devS.
4. Using the
keys, select one of these options: 2 or 3.
5. Press the key to save changes.
BypaSS uSer Id
This function allows the supervisor to selectively bypass the requirement to enter
a user ID before performing any test.
1. From the list of programmable parameters, choose BypaSS
using the
keys.
2. Press the key.
3. Select uSer Id using the
keys.
4. Use the
keys to select from two options: on or oFF.
5. Press the key.
maIn menu
Set parameterS
Qc parameterS
Note: If the
bypass is ON,
the User ID
requirement is
disabled and therefore
not required.
When the bypass
is OFF, the User ID
requirement operates
as normal and is
required.
maIn menu
Set parameterS
BypaSS