N-30 Rev F
Run Test
InStallatIon
N-30
Purpose
The Triage MeterPro Meter will prompt the user to install a Code Chip module when
required.
However, the InStall code chIp function may be used as an alternate method to transfer
information from a Code Chip module into the meter’s memory. The meter will direct the
user to install a Code Chip module if it is attempting to run a test and does not have the
data in its memory.
Types of Code Chip modules
There are five types of Code Chip modules
• Reagent Code Chip module included in each box
of test devices. Lot number begins with a W.
• QC Sample Code Chip module – included in each box
of controls. Lot number begins with a C.
• QC Device Code Chip module – included in the black
QC Device box (comes with the Triage MeterPro Meter) and
is labeled with the serial number of the QC Device.
• Supervisor Access Code Chip module – included with
each Triage
®
MeterPro
®
Meter. Lot Number begins with SPR.
• Software Upgrade Code Chip module – sent
separately when Triage MeterPro Meter software
upgrades are available.
cautIon: Running tests with the Supervisor Code Chip module installed disables
all QC Lockouts. Remove the Supervisor Cod e Chip module prior to running tests.
Install Code Chip
Software
Version Number