EasyManua.ls Logo

Biosite Triage MeterPro - Archive Last Test; Reagent Lots-QC

Biosite Triage MeterPro
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
O-24Rev F
operatIon
O-24
reagent lotSQc
A list of Reagent Code Chip modules in memory, along with corresponding expiration
dates for a QC Sample can be recalled from memory.
The list will contain: Lot Number, Panel Type, Expiration Date
• IftheQCexpirationdatehaspassed(expired),thedatewillbeinreversevideo
(light text on a dark background).
• IftheQCforthelothasfailed,thewordFaIled will appear in place of an
expiration date.
• IfaCodeChipmodulehasbeeninstalledforanewlot,butQChasyettobe
established, the words not run will appear in place of the expiration date.
• IfQCfrequencyissettonone,thereagentlotexpirationdatewillbedisplayed.
archIve laSt teSt
This function will save detailed information regarding the last performed test for analysis
by Biosite. Use this function when requested by Biosite Technical Services.
This function should be used when a very unusual test result is received and when the
laboratory wishes to have the meter evaluated further. archIve laSt teSt will only be able
to save data from the most recent test run.
The Reagent Lots - QC Screen
maIn menu
recall reSultS
Qc reSultS
reagentS lotS Qc
archIve laSt teSt