O-2Rev F
Run Test
operatIon
O-2
Purpose
Run Test is used for the analysis of a test.
Options
The Run test function allows you to test:
• QCDevice
• QCSample
• PatientSample
• Misc.TestSample
In order to run a patient sample, the meter requires:
1. A valid User ID has been entered (unless this feature is bypassed or a Supervisor
Code Chip module is installed). See page N-21.
2. A QC Device has passed within the specified time period. See page O-3.
3. A QC Sample for the test device lot has passed within the specified time
period. See page O-5.
If any of these conditions do not exist, the meter will not continue and will prompt the user
to perform the required steps.
Important: The Triage MeterPro Meter will alert the user if QC has not been
performed for the device lot being tested.
Note: The
Supervisor
Code Chip module
provides access to
functions that may
override these
Lockouts.
Note: If the
Triage MeterPro
Meter’s internal quality
control (QC) results are
unacceptable,
exclamationpoints(!)
will be displayed on the
screen in the place
of the patient results.
To obtain results, repeat
the test using a new
device.