O-3 Rev F
operatIon
O-3
maIn menu
run teSt
Qc devIce
The QC Device should be run on each day of patient testing.
To run the QC Device test:
1. Press the key to power The Triage MeterPro Meter on.
2. Select run teSt using the
keys.
3. Press the key.
4. Type in your identification number (User ID). To correct, press the key to
clear the entire ID or select the incorrect character by using the
keys. Then type in the correct ID.
5. Press the key
6. Select Qc devIce using the
keys.
7. Press the key.
8. Gently insert the QC Device into the Triage MeterPro Meter until you feel
the QC Device catch on the pin and hear an audible ‘click.’
9. Press the key.
Note: If the
User ID bypass
is ON, the meter will
not ask you to enter a
User ID.
Menu screen as it looks when operator
selects QC Device.
“EXP” indicates that
the QC Device re-test
frequency has expired.
The Main Menu
Note: The meter
will prompt the
user to install the QC
Device Code Chip
module if the QC
Device has not been
run before.