EasyManua.ls Logo

Biosite Triage MeterPro - Page 45

Biosite Triage MeterPro
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
O-4Rev F
operatIon
O-4
10. The Triage MeterPro Meter pulls in the QC Device and scans it. The test
device may partially move in and out of the meter several times.
11. When the test is complete, the meter will beep, eject the device and display the
results on the meter’s screen.
12. Press the key to make a printed copy of the results.
Important InStructIonS For all teStS:
a. Gently insert the test device into the Triage MeterPro Meter until you feel the
device catch on the pin. You will hear an audible ‘click’ that tells you
the test device has been inserted properly.
b. Press the key to start the test.
c. The Triage MeterPro Meter pulls in the test device and scans it.
d. When the test is complete, the meter will beep and display the results
on the meter’s screen.
e. Press the key to make a printed copy of the results.
Important InStructIonS For Qc devIce teStS:
• Dust,lint,fibersandothersmallparticlesmayinterferewiththeQCDevice.
Keep the QC Device free of contaminants.
• TheQCDeviceislight-sensitiveandshouldbestoredinitsblack
opaque case when not in use.
• IftheQCDevicetestsfail,wipetheQCDevicecleanwithalint
free cloth to remove any oils, dust, fibers, or fingerprints. Do not apply
any liquid to the QC Device. After cleaning the device, repeat
the QC Device test.
• IftheQCDevicetestfailsafteryouhavecleanedthedevice,call
Biosite Incorporated Technical Services at 1-877-441-7440.
• Note:theQCDevicedoesnotexpire.
maIn menu
run teSt
Afterthetest
device has been
inserted, do not
push the device in
further or attempt to
pull it out. The device
may be ejected by
returning to the Main
Menu and pressing the
EXIT key.

Related product manuals