N-22 Rev F
InStallatIon
N-22
uSer Id
Up to 600 User ID’s with 1 – 16 characters digits may be programmed into the meter.
The User ID may alternately be entered using the external Bar Code Scanner (optional).
An additional security feature of the User ID is the partial masking of the ID on displays
and printouts. This prevents unauthorized individuals from detecting and using an existing
User ID. When the Supervisor Code Chip module is installed in the meter, this feature is
overridden and all User ID characters are displayed.
new uSer Id
1. From the list of programmable parameters, choose uSer Id
using the
keys.
2. Press the key.
3. Select new uSer Id using the
keys.
4. Press the key.
5. Type in a User ID with a number of characters within mIn characterS uSer Id and
Max Characters User ID. To correct, press the key to clear the entire ID, or
select the incorrect character by using the keys. Then type in the
correct number.
6. Press the key.
7. On the next screen, use the
keys to select an expiration date to the
right of uSer Id StatuS: 6 monthS, 12 monthS, valId or expIred.
8. Press the key to save changes.
The Triage module MeterPro Meter recognizes leading 0’s as a unique character.
Therefore, ‘4341’ will be recognized separately from ‘04341’.
Using the barcode scanner or alpha mode (shift key) will allow the Triage module MeterPro
Meter to accept alpha characters and the following symbols: # ( ) * - . / \
maIn menu
Set parameterS
uSer Id
Note: USERID
9999999999
(10-9’s) is a
default setting
displayed when the
USERIDBYPASS is on
or when a test
is run with the
Supervisor Code Chip
module installed.
Note: If you
type in a User
ID that has already
been assigned, the
message on the
screen states:
USER###ALREADY
EXISTS.
Press the ENTER key
to go back and type