N-7Rev F
InStallatIon
N-7
For more details
aboutRangesSettings
see page N-25.
For more details about
Test Settings
see page
N-26.
Note: The lowest
value of the
RangeorThreshold
may be disabled for
sometestpanels.Refer
to the appropriate
Product Insert for
specific information.
8. Change Reference Ranges / Thresholds for Tests:
a. Select Set parameterS using the
keys and press the key.
b. Select rangeS using the
keys and press the key.
c. Select the test panel type using the
keys.
d. Select the analyte and level using the
keys. Type in the new value for
the hIgh or low cutoff as appropriate.
e. Press the key to save changes.
9. Change Test Settings:
a. Select set paraMeters using the
keys and press the key.
b. Select test settIngs using the
keys and press the key.
c. Select the test panel type using the
keys.
d. Select the analyte using the
keys.
e. Select the test setting (ACTIV, INACT, USR A, or USR I) using the
keys.
f. Press the
key to save changes.
10. Change QC Parameters:
a. Select Set parameterS using the
keys and press the key.
b. Select Qc parameterS using the
keys and press the key.
c. Select the desired setting using the
keys and change
the value to the right of the setting by using the
keys.
Setting Default Options New Setting
Qc devIce FreQ. Daily None, 8HR, Daily,
Weekly, Monthly
Qc Sample FreQ. Monthly None, 8HR, Daily,
Weekly, Monthly
numBer oF controlS 2 1 or 2
num Quant Std devS 3 2 or 3
d. Press the key to save changes to QC Parameters.
Note: The ability
to change test
settings may be
disabled for some tests
inspecificpanels.Refer
to the appropriate
Product Insert for
specific information.
For more details
about QC Parameters
Settings see page
N-28.