N-8 Rev F
InStallatIon
N-8
11. Change Bypass settings:
a. Select Set parameterS using the
keys and press the key.
b. Select BypaSS using the
keys and press the key.
c. For uSer Id select on or oFF using the
keys.
d. Press the key to save changes to the Bypass setting.
12. Prior to commencing patient testing, remove the Supervisor Code Chip
module
from the meter, return it to the storage box and place in a safe place for future
use.
The meter is now ready to run QC Tests in preparation for Patient Testing.
QC Testing should be performed in the following manner:
1. Run QC Device (see page O-3).
2. Run QC Sample for each lot of devices (see page O-5).
3. If appropriate, Run Calibration Verification Set as a Misc Test sample for each test
panel type to be used. (Refer to the applicable Procedure Manual and Calibration
Verification Set Package Insert for detailed instructions).
Location
The Triage MeterPro Meter should be placed on a dry, clean, flat, horizontal surface away
from direct sunlight.
Power
The Triage MeterPro Meter requires 6-volts DC supplied by either AC or DC. The meter
will run approximately 100 tests on batteries. It is recommended to use batteries as a
backup
in the event of a power outage.
cautIon: Use only the power supply provided with the meter, failure to do
so may cause permanent damage to the meter.
For more details about
Bypass Settings see
page N-28
Note: If the
bypass is
on
, the
user ID feature is
disabled and anyone
may run a test.
If the bypass is
oFF
,
the user ID feature
is active and only
authorized users may
run a test.
To Install User ID’s,
see page N-22.
Note: The Triage
MeterPro Meter
will automatically
turn off if left unused
for 2 hours, unless
programmed
otherwise by the
supervisor.Optional
settings allow
automated turnoff
after
1
/2 hour, 1 hour,
2 hours, 4 hours or
NONE.