N-9Rev F
InStallatIon
N-9
Self-Test
The Triage MeterPro Meter checks the system to verify it is operating properly by running
self-tests when powered on and prior to each test.
• Power On – includes laser operation, internal standard range, battery power,
and software verification.
• Each Test – verifies the laser operation, internal standard range and battery power.
Code Chip modules
Meter data is updated via a disposable Code Chip module. The Code Chip module
contains microchip circuitry embedded into a plastic housing. When inserted into the
meter Code Chip module port and activated, the information is transferred into the Triage
MeterPro Meter’s memory. A Code Chip module typically needs to be installed only once
and remains in the meter memory. A Code Chip module does not need to remain in the
meter while performing tests.
There are five types of Code Chip modules which can be distinguished by shape or color.
• Reagent Test Code Chip module– included in each box of test devices; contains
calibration, expiration date and other data about the device lot. (Color varies by test type.)
• QC Sample Code Chip module – included in each box of QC Sample; contains
expiration date and other data about the QC Sample lot including acceptable range.
(Color varies by test type.)
• QC Device Code Chip module – included in the black QC Device box (comes
with the Triage
MeterPro Meter); contains data about how the meter should read its
matching
QC Device. (Dark gray.)
• Supervisor Access Code Chip module – included with each Triage
MeterPro Meter;
contains code to activate Supervisor functions. (Light gray.)
• Software Upgrade Code Chip module – provided when Triage MeterPro Meter
software upgrades are available; contains new software for the meter.
Software Upgrade Code Chip module
Standard Code Chip module