N-27Rev F
InStallatIon
N-27
teSt SettIngS
1. From the list of programmable parameters, choose teSt SettIngS
using the
keys.
2. Press the key.
3. Select the test panel type using the
keys.
4. Select the analyte using the
keys.
5. Select the desired mode using the
keys.
• actIve (actIv on Screen) – The test will run on every patient specimen.
• InactIve (Inact on Screen) – The test will never run on a patient specimen and is
dropped from all displays and printouts.
• uSer InactIve (uSr I on Screen) At the time of patient testing, the user has the
ability to select the test.
• uSer actIve (uSr a on Screen) At the time of patient testing, the user has the
ability to deselect the test.
6. Press the key to save changes.
Qc teStS wIth teSt SettIngS actIvated
Any test which the Supervisor has set to actIve, uSer actIve or uSer InactIve will run on
the QC Sample panel. The meter will display and print the test’s QC sample result, and
update the QC Sample timer for the test. If the Supervisor has set the test to InactIve, the
test will have the QC Sample turned OFF. The QC Sample result won’t show for that test
and the QC Sample timer won’t be updated for it.
If the Supervisor changes the test from InactIve to actIve, to uSer actIve or to uSer InactIve
that test may be out of QC Sample date. As always for a test with an expired QC Sample,
a passing QC Sample test will need to run before a patient sample can report a result for
that test.
maIn menu
Set parameterS
Note: When
the analyte is
deselected, the value
is not calculated nor
stored.