O-9c Rev F
operatIon
O-9c
11. Prepare the test sample and perform the test in accordance to the instructions
provided in the package of test devices you are using.
Important InStructIonS For all teStS:
a. Gently insert the test device into the Triage Meter until you feel the
device catch on the pin. You will hear an audible ‘click’ that tells you
the test device has been inserted properly.
b. Press the key to start the test.
c. The Triage Meter pulls in the test device and scans it.
d. When the test is complete, the meter will beep and display the results
on the meter’s screen.
e. Press the key to make a printed copy of the results.
Important InStructIonS For mISc. teStS:
a. If the test device’s Internal Quality Control (QC) results are unacceptable, the
results for the affected analyte(s) will not appear on the screen (instead will
be flagged with a ! symbol).
maIn menu
run teSt
Afterthetest
device has been
inserted, do not
push the device in
further or attempt to
pull it out. The device
may be ejected by
returning to the Main
Menu and pressing
EXIT.