EasyManua.ls Logo

BIRD FlightHawk - Chapter 3 Calibration; Calibrating the FlightHawk-AV Analyzer

BIRD FlightHawk
130 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
28
Chapter 3 Calibration
Calibrating the FlightHawk-AV Analyzer
For best results, set the frequency and calibrate the Bird Site Analyzer
immediately before taking measurements.
Note: Use a precision open, short, load (OSL) calibration standard or
“Calibration Combo” for accurate measurement results.
Note: When using an extension cable, a phase stable cable is needed
to ensure performance and accuracy.
Note: Changing frequency settings will automatically turn calibration
off. Always set the frequency before calibrating the unit.
1. Press the Calibrate icon. See
Figure 27
.
2. Connect the Cal Combo “Open” connector to the unit antenna port.
3. Press the Open softkey.
4. Wait for the unit to sweep and a Green Check to display.
5. Connect the Cal Combo “Short” connector to the unit antenna port.
6. Press the Short softkey
7. Wait for the unit to sweep and a Green Check to display.
8. Connect the Cal Combo “Load” connector to the unit antenna port.
9. Press the Load softkey
10. Wait for the unit to sweep and a Green Check to display.
Note: During calibration, the color of the icon display is yellow.
Note: After calibration, the display will resemble
Figure 27
and
Correction will be set to On.
Figure 27 Calibration Display Screen


 




Table of Contents

Related product manuals