10 Technical specifications
118 Operating Manual ARC 350 900-351_IFU_V2.0.0_20221-S0-20150520-EN
Bipolar Coagulation – Bipolar Resection
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Bipolar Coagulation Bipolar
Resection"
= 175 W
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Bipolar Coagulation Bipolar
Resection"
= 350 W
Diagram of power output P [W] as a
function of the setting " Bipolar
Coagulation Bipolar Resection "
Rated load resistance = 25
HF output voltage U [Vp] for the setting
" Bipolar Coagulation Bipolar Resection
" (idle mode)
= 190 Vp