5 Operation
50 Operating Manual ARC 350 900-351_IFU_V2.0.0_20221-S0-20150520-EN
5.7.11. SimCoag
This mode is used for simultaneous preparation. Two monopolar socket
connectors can be activated at the same time to enable the simultaneous use of
two manually switched instruments. Both handles can be switched on and off
independently.
The current type changes with the effect setting:
Effect 1: Forced cutting
Effect 2: Forced mixed
Effect 3: Spray
The output power setting applies to both output sockets, and the power
is distributed according to the tissue structure.
Application areas
Simultaneous coagulation and preparation, e.g. for cardiac or mammary surgery
Suitable instruments
Ball electrodes
Knife electrodes
Spatula electrodes
5.7.12. Gastro Coag (optional)
This mode is used in gastroenterology with contact coagulation for the coagulation
of small areas.
Application areas
After bleeding associated with polypectomy or papillotomy.
Suitable instruments
Polypectomy snares
Papillotome
This function is available if the device has the options argon / GastroCut
(900-391).
5.7.13. Laparoscopy
This mode is used in laparoscopy and arthroscopy for monopolar coagulation.
Application areas
Laparoscopy, arthroscopy
Suitable instruments
Arthroscopy electrodes
Laparoscopy electrodes