5 Operation
900-351_IFU_V2.0.0_20221-S0-20150520-EN Operating Manual ARC 350 55
5.9.7. Bipolar scissors
This mode is used with bipolar scissors. It can be used for coagulation before or
during cutting, point coagulation, coagulation of cuts and surface coagulation.
Application areas
Preparation, coagulation and cutting of tissue
Suitable instruments
Bipolar scissors
Bipolar scissors should only be operated with the current type bipolar
scissors cutting or bipolar scissors coagulation.
5.9.8. Laparoscopy
This mode is used for coagulation in combination with bipolar laparoscopic
instruments.
Application areas
Laparoscopic coagulation
Suitable instruments
Laparoscopic instruments
5.9.9. Bipolar resection (optional)
This mode is used for bipolar haemostasis in gynaecology and in urology for
cutting under saline.
Use NaCl as irrigation fluid.
Application areas
Hysteroscopy, transurethral prostate resection (TUR-P), surgical treatment of
bladder tumours (TUR-B), vaporization of prostate tissue (TUR-VAP)
Suitable instruments
Resectoscope
Resection snare
Rollerblade electrode
Make sure that the instrument has contact with the tissue while activating
bipolar coagulation to avoid an unintended heating of the irrigation fluid.
This function is available if the device has the option Bipolar Resection
(900-395)