5 Operation
900-351_IFU_V2.0.0_20221-S0-20150520-EN Operating Manual ARC 350 49
5.7.9. Resection
This mode is used for monopolar hemostasis in gynaecology and urology.
Use non-conductive irrigation fluids.
Application areas
Hysteroscopy, transurethral prostate resection (TUR-P), surgical treatment of
bladder tumors (TUR-B), vaporization of prostate tissue (TUR-VAP).
Suitable instruments
Resectoscope (monopolar)
Resection snare
Rollerblade electrode
5.7.10. Cardiac Mammary
This mode is used in mammary surgery and cardiac surgery. It produces forced
coagulation.
Application areas
Mammary surgery and cardiac surgery
Suitable instruments
Knife electrodes
5.7.10. Cardiac Thorax
This mode is used in thoracic surgery. It produces forced coagulation.
Application areas
Thoracic surgery
Suitable instruments
Knife electrodes