5 Operation
48 Operating Manual ARC 350 900-351_IFU_V2.0.0_20221-S0-20150520-EN
5.7.7. Argon flexible (optional)
This mode is used for argon-assisted electrosurgery in conjunction with the ARC
PLUS accessory device.
This is the current type Spray.
For argon-assisted coagulation, flexible probes are used in combination with
endoscopes.
Application areas
Gastroenterology, homogeneous surface coagulation
Suitable instruments
Flexible argon probes
This function is available if the device has the options argon / GastroCut
(900-391).
5.7.8. Argon flex. pulse (optional)
This mode is used for argon-assisted electrosurgery in conjunction with the ARC
PLUS accessory device.
This is the current type Spray.
For argon-assisted coagulation, flexible probes are used in combination with
endoscopes.
The puls frequency changes with the effect setting. The higher the effect level, the
faster the pulse sequence.
Application areas
Gastroenterology, homogeneous surface coagulation
Suitable instruments
Flexible argon probes
This function is available if the device has the options argon / GastroCut
(900-391).