6 Detecting and correcting faults
68 Operating Manual ARC 350 900-351_IFU_V2.0.0_20221-S0-20150520-EN
Mixed operation of argon bottles with and without an electric bottle
pressure gauge is not recommended
Connect two identical pressure reducers.
Please check if the instrument is free of adhesions, and purge it with
argon.
If repeated purging does not solve the problem, the instrument and
cable must be replaced.
The argon flow settings at ARC 350 are invalid
The filling level of the argon bottle is low.
Please make sure that a replacement is available.
You can connect two argon bottles. The unit shifts automatically to
the second bottle.
The argon bottle is empty.
Connect a replacement bottle to enable activation.
You can connect two argon bottles. The unit shifts automatically to
the second bottle
The argon bottle is empty. The unit has shifted automatically to the
replacement bottle.
Please make sure that a replacement is available
The annual Technical Safety Inspection (TSI) for ARC PLUS is due.
Plug'n Cut COMFORT
Notice
The lifetime of the instrument is ending soon.
Please order a replacement in good time.
Any use of the instrument beyond its lifetime is not covered by
warranty. Please contact your BOWA dealer in good time to
purchase a new instrument.
Plug'n Cut COMFORT
Warning
The maximum lifetime of the instrument has been reached. Any
further use is not covered by warranty. The maximum service
lifetime of the instruments must not be exceeded, in order to
guarantee safe usage. Any further use is at the user's risk.
Plug'n Cut COMFORT
Notice
You have connected a BOWA RFID instrument. X applications
remain available. The maximum lifetime of the instruments must not
be exceeded, in order to guarantee safe usage. Any further use is at
the user's risk.
You can save up to six of your preferred programs on your Dr.
Dongle and run them in a flexible way. With the "Load" button you
can call a selected program. You can use the "Save" button to enter
an explanatory text for the selected program.
Plug'n Cut COMFORT
Warning
A software update is necessary to use Plug'n Cut COMFORT with
this instrument. Only carry out manual settings at this instrument.
Please contact the Technical Support. Internet: www.bowa.de
Telephone: +49707260020
If this message appears again, please contact the Technical
Support. Internet: www.bowa.de Telephone: +49707260020
Internal Errors have a number next to the description.
Please advise the Technical Service of this number.