EasyManua.ls Logo

Bowa ARC 350 - Page 80

Bowa ARC 350
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 Technical specifications
80 Operating Manual ARC 350 900-351_IFU_V2.0.0_20221-S0-20150520-EN
Description
CCS
ARC
CONTROL
Form of HF
voltage
Max. power output
Peak
voltage
Default values
Effect
Power range
Effect
Max. Watt
Modes Bipolar Coagulation
Micro forceps
sinusoidal
constant
-
0.1 W - 20 W
150 Vp
---
10
Forceps
forced
sinusoidal
modulated
-
1 W - 100 W
550 Vp
---
70
LIGATION
sinusoidal
modulated
-
200 W
190 Vp
---
---
TissueSeal
PLUS
sinusoidal
modulated
-
200 W
190 Vp
---
---
Bipolar
scissors
sinusoidal
constant
-
1 W - 120 W
200 Vp
---
60
Laparoscopy
sinusoidal
constant
-
1 W - 120 W
150 Vp
---
50
Bipolar
resection
sinusoidal
constant
-
1 W - 350 W
190 Vp
---
200
The max. values are not necessarily created at rated load resistance.
The HF power is subject to a tolerance limit of ± 20%.