10 Technical specifications
84 Operating Manual ARC 350 900-351_IFU_V2.0.0_20221-S0-20150520-EN
Monopolar Cutting Argon
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Monopolar Cutting Argon"
= 150 W
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Monopolar Cutting Argon"
= 300 W
Diagram of power output P [W] as a
function of the setting "Monopolar
Cutting Argon"
Rated load resistance= 500
Effect
U (Vp)
1
400
2
450
3
560
4
650
5
650
6
700
7
700
8
700
9
750
Table of HF output voltage U [Vp] as a
function of the setting "Monopolar
Cutting Argon" (idle mode)