EasyManua.ls Logo

Bowa ARC 350 - Page 90

Bowa ARC 350
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 Technical specifications
90 Operating Manual ARC 350 900-351_IFU_V2.0.0_20221-S0-20150520-EN
Monopolar Cutting GastroLOOP 3
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Monopolar Cutting
GastroLOOP 3"
coag phase
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Monopolar Cutting
GastroLOOP 3"
cut phase
Effect
P (W)
coag phase
P (W)
cut phase
1
35
400
2
35
400
3
35
400
4
35
400
5
35
400
6
35
400
7
35
400
8
35
400
9
35
400
Table of power output P [W] as a
function of the setting "Monopolar
Cutting GastroLOOP 3"
Rated load resistance= 500
Effect
U (Vp)
1
750
2
750
3
750
4
750
5
750
6
750
7
750
8
750
9
750
Table of HF output voltage U [Vp] as a
function of the setting "Monopolar
Cutting GastroLOOP 3"
(idle mode)

Table of Contents

Other manuals for Bowa ARC 350

Related product manuals