10 Technical specifications
98 Operating Manual ARC 350 900-351_IFU_V2.0.0_20221-S0-20150520-EN
Monopolar Coagulation Spray
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Monopolar Coagulation
Spray"
= 60 W
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Monopolar Coagulation
Spray"
= 120 W
Diagram of power output P [W] as a
function of the setting "Monopolar
Coagulation Spray"
Rated load resistance = 500
Effect
U (Vp)
1
3000
2
3800
3
4600
4
5000
Table of HF output voltage U [Vp] as a
function of the setting "Monopolar
Coagulation Spray" (idle mode)