Disassembly / Assembly
3 - 56 Infusomat® Space 6.0
EN
Check List for Checks after Repair
Unit: Infusomat® Space
Article No. Unit No.
Manufacturer: B. Braun Melsungen AG
Visual Inspection Electrical Safety
According to IEC 60601-1 or IEC 62353
Functional Inspection
❒ Cleanliness, completeness, damage and
fault affecting safety
, damage to and
readability of the labels.
Special:
❒ Co
nnectors "P2" and "P3"
❒ Safety cla
mp
❒ Pressure sensor membra
ne, down-
stream
❒ Pressure sensor membra
ne, up-
stream
❒ Pu
mp membrane
❒ Pum
p slide
(All 12 slides are engaged)
❒ Co
ding of the pump slide guide
❒ Pump connecting ro
d
❒ Air inline
sensor (free of grease or
ultrasound gel, free of cracks)
❒ Drop sensor (lens)
The patient and housing leakage current of
the Infusomat® Space is caused exclusively
by the operating voltage supply (Power Sup-
ply SP or SpaceStation).
The Technical Safety Checks of the power
supply SP
(drawing No. M001321005F04) or
of the SpaceStation (drawing No.
M690000046F04) serve to check whether
both limit values are met.
❒ Locking with
second unit
❒ Ba
ttery compartment cover
Switch on unit with power supply:
❒ Self-test
❒ Ind
icator lamps (LEDs)
❒ Audible alarm
❒ Visual alarm
❒ Status displa
y
❒ Lighting
Operation
❒ Opening and closing mech
anism of the
operating unit
❒ Slid
e clamp releases release lever of the
safety clamp. The Infusomat® Space
Line is pinched.
❒ Buttons on the op
erating unit
❒ Infusion
❒ Staff call
❒ T
rigger bolus at the device
❒ T
rigger bolus by pressing the PCA but-
ton
❒ Slid
e clamp is completely pulled onto
the Infusomat® Space Line when the
operating unit is opened.
❒ The operating lever of the safety clamp
can be pressed down evenly against the
spring tension and engages. The yellow
caution sign flashes.
(Part 1 of 3)