Disassembly / Assembly
3
Infusomat® Space 6.0 3 - 57EN
Pressure switch-off, electronic:
(see TSC for limit values)
Delivery rate: 250 ml/h
Testing without drop sensor
Infusomat® Space Line, standard type
Upstream
Alar
m
Downstream
Pressure stage 2
Pressure stage 5
Pressure stage 9
Pressure limitation, mechanical
(see TSC for limit values)
Pm
ax
Pm
in
Safety clamp
(see TSC for limit values)
Pm
in
Flow accuracy
(see TSC for limit values)
Room temperature 20 … 25°C, delivery rate: 250 ml/h,
measuring quantity: 25 ml
Deviation within tolerance
Drop sensor (optional)
Delivery rate: 250 ml/h
Drop sensor alarm passage
Drop sensor alarm closure
Visual Inspection Electrical Safety
According to IEC 60601-1 or IEC 62353
Functional Inspection
(Part 2 of 3)