Disassembly / Assembly
3
3 - 58 Infusomat® Space 6.0
EN
Air inline sensor
(see TSC for limit values)
Infusomat® Space Line, standard type
Water value
Difference of
temperature sensors
Air
alarm
Air
value
Note: If the water or air value is not
reached immediately, these values have to
be read again after 120 seconds.
Switch on unit without power supply
Self-test
Magnetic function of the battery co
m-
partment cover
Charg
ing level of the battery module
Remove battery module
Alarm fo
r at least 3 minutes.
Visual Inspection Electrical Safety
According to IEC 60601-1 or IEC 62353
Functional Inspection
(Part 3 of 3)
Date: Name: Signature: