Infusomat® Space 6.0 4 - 1EN
4 - Servicing the Unit
Cleaning and Disinfecting
Notes
WARNING
!
Liquid may penetrate live parts of the unit.
DANGER OF ELECTRIC SHOCK.
- Before disinfecting the pump, always disconnect the pump
from Infusomat® Space the patient, switch off the device and
disconnect from power and other devices (e.g. staff call).
- To reduce the likelihood of moisture ingress into the electrical
connectors, the P2 connector of a power supply or connec-
tion cable may be used to cover the connections during
cl
eaning operations. Ensure that any connectors used to cov-
er are not connected to a wall outlet
or other voltage source.
- Do not touch any parts of the wiri
ng or the tube pump with
sharp objects.
- We recommend using the cleaning and disinfecting agents
M
eliseptol® Wipes sensitive or Meliseptol® Foam pure from
B. Braun for cleaning and disinfecting.
- In the case of severe soiling that cannot be removed by the
above method, please contact the hospital's technical de-
partment.
- The current internal hygiene measure
s of the hospital apply.
Housing
Clean and disinfect all surfaces of the Infusomat® Space.
Start by cleaning the display with a dampened nonwoven cloth.
Cle
an using a damp and NOT a soaking cloth. Dispose of the cloth
after use.
We recommend: Meliseptol® Wipes
sensitive or Meliseptol®
Foam pure.
You can use Meliseptol® Foam pure to clean the flex cable of the
Infusomat® Sp
ace. Apply with a cotton swab..
Make sure that no liquid is allowed to enter the contacts or ven-
tilation slots. Allow the device to dry after cleaning/disinfecting
(for a
t least 1 minute).
When they are dry, insert the cleaned
/disinfected Space infusion
pumps back into the holder or SpaceStation (see cleaning plan for
SpaceStation) and protect from dirt until they are next used.
Fig.: 4 - 1
4 - 4

Other manuals for Braun Infusomat Space

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Infusomat Space and is the answer not in the manual?

Braun Infusomat Space Specifications

General IconGeneral
TypeInfusion Pump
Power SupplyAC/DC adapter, rechargeable battery
DisplayColor graphic display
AlarmsAudible and visual alarms
Flow Rate Accuracy±5%
Occlusion PressureAdjustable
VTBI Range0.1 – 9999 ml
KVO Rate0.1 – 5 ml/h (adjustable)
BatteryLithium-ion
Operating ModesContinuous
CertificationsCE, ISO 13485, IEC 60601-1