Procedural Instructions on the TSC
6 - 4 Infusomat® Space 6.0
EN
2. Battery check
Perform the battery service to discharge the battery until the
battery pre-alarm is issued (no bar displayed
for battery dis-
charge level (Fig.: 6 - 4), but it must not flash).
Open and close the operating unit 5x.
- Replace the battery if the battery alarm sounds while you
o
pen or close the operating unit.
3. Switch on the device with the power supply connected
and
check the following points:
- Self-test
Display on the LC display
-
Indicator lamps
LEDs (yellow, green, blue)
*
light up for a short moment
LED changes from green to red
†
- Audible alarm
A low and then a high sound
- Visual alarm
Red LED lights up for a short moment
- Self-test
Display on the LC display
- Status display
Battery capacity, mains operation
- Lighting
LC display lighting, keyboard lighting
4. Operation
a) Open operating unit.
Yellow warning light beside the safety clamp flashes.
b) Insert Infusomat® Space Line.
Slide clamp releases release lever of the safety clamp.
The
Infusomat® Space Line is pinched.
c) Close operating unit.
*
As of software version J: yellow, red, blue
†
As of software version J: from red to green