Procedural Instructions on the TSC
6
Infusomat® Space 6.0 6 - 3EN
Functional Inspection Infusomat® Space
Mechanical Inspection
1. Fit the unit to be tested on top
of another Space device and
check the proper functioning of the lock.
2. Fit the unit to be tested under another Sp
ace device and
check the proper functioning of the lock.
3. Check the battery compartment cover lock for proper opera-
tion.
Functional Check
1. Determine age of battery
Open battery compartment co
ver and remove battery.
Read batch number on the battery nameplate and determine
a
ge of the battery:
a) Please see Fig.: 6 - 2 for NiMH batteries.
b) Please see Fig.: 6 - 3 for Li-ion batteries.
Note
If the battery is more than 8 years old, recommend changing the
battery or conducting a battery check.
Fig.: 6 - 2 NiMH batteries Fig.: 6 - 3 Li-ion batteries
Letter: Year of manufacture
according to table left
Two digits: Calendar week
Example:
Manufactured in week 42/2008
D 2003
E 2004
F 2005
G 2006
H 2007
I 2008
...
N2013
First two digits: Calendar week
Second two digits: Year of manufacture
Example:
Manufactured in week 41/2011