Disassembly / Assembly
3 - 40 Infusomat® Space 6.0
EN
3.14 Installation / Assembly
Assembly or installation of the modules and subsystems is per-
formed in reverse order of disassembly. Special steps to be ob-
served are described hereafter in detail.
Only new cover caps are to be used.
Special Screws
Special screws for plastic housings are used in
this unit. These are
special screws for plastic housings. The screws are not self-cut-
ting but produce a thread in the plastic of the housing through
d
eformation when fitted in for the first time.
If the beginning of the thread is not
engaged when the screw is
fitted, a new thread is produced. The old thread is destroyed. As a
result, the strength of the screw connection is longer guaranteed.
Screw in the special screws for plastic housings
as follows:
1. Position the special screw on the thread.
2. Rotate screw anti-clockwise (loosen) until a faint click can be
heard. This click is produced
when the screw thread drops
into the existing thread.
3. Screw in screw and tighten carefully.
Zero Force Insertion Connector
Make sure that the ribbon cable is centered between the plug
guides when the zero force insertion connector is locked. Check
the lock and the ribbon cable for correct fit before any further in
-
stallation.