Disassembly / Assembly
3
Infusomat® Space 6.0 3 - 41EN
Processor PCB
1. The adjusting lug of the bottom inner
frame must be posi-
tioned in the processor PCB boring.
When the processor PCB is replaced all data of the pump except
fo
r the calibration data was probably saved on a PC(see „Prepara-
tions for Exchanging the Processor PCB“ pg
. 3 - 2). Carry out
the following steps to transmit the data back to the device (see
„Rework when Replacing the Processor PCB“ pg. 3 - 46).
Operating Unit
1. Insert the hinge spring with its long side in the hinge cover
o
pening, spring, and with its angled side in the rear panel
bore hole of the operating unit.
2. Please pay attention to the ground connection on the key-
board PCB when mounting the rear panel of the operating
un
it.
3. The grooves of the hinge covers are pointing
to the operating
unit rear panel.
4. Push the operating unit connection cable into
the processor
PCB connector.
5. Lock the connector.
6. The ribbon cable cover of the operating unit must be laid be-
tween the bottom inner frame and the operating unit and
o
ver the connector.
7. Check the seal washers for correct position and completeness
o
nce again after assembly.
Fig.: 3 - 43