Disassembly / Assembly
Infusomat® Space 6.0 3 - 53EN
6. Unpack the zip file with the new tube data (available on re-
quest) into the following directory
C:\B.Braun Space\HiBaSeD_8.0.1_1\dat\Infusomat.
Alternatively: Copy the unpacked folders “defa
ult” and “Dis-
posables” to that same directory.
7. Start “HiBaSeD 8.0.1_1_neue Schlauc
hdaten” by double-
clicking on the link.
8. Ontab “Service” > “Calibration“, press butto
n “Modify”.
9. Only choose “Disposable article” fro
m the data selection.
10. Start calibration with “OK”.
The new tube data are imported.
11. In menu
Line selection
check whether the new disposable ar-
ticle table exists in the device.