Host Commands Polycold Cryochiller
System Configuration Commands Installation and Operation Manual
Brooks Automation
7-10
214072 Revision B
VENDOR_ID Get the DeviceNet Vendor ID
IP_ADDR IP Address
DEFAULT_GATEWAY Default Gateway
Purpose: Returns identifier string indicating the vendor ID.
Only returns a value if DeviceNet config is installed
Parameter(s): ?
Return Value: [A][cccccccc]
Default Value: N/A
Security Level: Get: 1 Set: N/A
Example: “VENDOR_ID,?”
Interface RS-232, Ethernet
Purpose: Get/set the IP Address for the system.
Parameter(s): ? or [nnn.nnn.nnn.nnn]
[nnn.nnn.nnn.nnn] is the IP address
Return Value: [A] or [A][nnn.nnn.nnn.nnn]
Default Value: 0.0.0.0
Security Level: Get: 1 Set: 3
Example: “IP_ADDR,192.168.1.100” Set the IP Address for host port
Interface: RS-232, Ethernet
Purpose: Get/set the default network gateway for the system.
Parameter(s): ? or [nnn.nnn.nnn.nnn]
[nnn.nnn.nnn.nnn] is the Gateway address
Return Value: [A] or [A][nnn.nnn.nnn.nnn]
Default Value: 0.0.0.0
Security Level: Get: 1 Set: 3
Example: “DEFAULT_GATEWAY,192.168.1.1” Set the default gateway for host port
Interface RS-232, Ethernet