Polycold Cryochiller Host Commands
Installation and Operation Manual System Sensor Readings
Brooks Automation
214072 Revision B
7-23
E_MGMT_ENABLE Energy Management Enable
System Sensor Readings
SENSOR_SNAPSHOT Sensor Data Snapshot
Purpose: Get/set the enable state of energy management. (also referred to as
power management)
Monitor Data:
Parameter(s): ? or [n]
[n] is
1 = enabled
0 = disabled
Return Value: [A] or [A][n]
Data Type: Integer
Default Value: 1
Min: 0 Max: 1
Units: N/A
Security Level: Get: 1 Set: 3
Example: “E_MGMT_ENABLE,?” Get the enable state of energy management
Interface RS-232, Ethernet
Purpose: Get a snapshot of the current sensor data.
Parameter(s): ?
Return Value [A][nnnnnnnn1],[nnnnnnnn2],[nnn.n]….
Param 1 is bit field for if sensors values are valid (valid = 1) hex encoded
bit 0 – 20: T1 – T21
bit 21: Voltage transformer
bit 22: Not Used (0)
bit 23: Not Used (0)
bit 24: Not Used (0)
bit 25: Not Used (0)
bit 26: Not Used (0)
bit 27: Discharge Pressure Transducer
bit 28: Suction Pressure Transducer
bit 29: Not Used (0)
bit 30: Not Used (0)
bit 31: Not Used (0)