Brooks Polycold MaxCool Series - Alarm_Status; Current Alarm Status

Brooks Polycold MaxCool Series
250 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Host Commands Polycold Cryochiller
System Operational State Commands Installation and Operation Manual
Brooks Automation
7-18
214072 Revision B
ALARM_STATUS Current Alarm Status
Return Value bit 11: C2 OUTLET SENSOR (TC12)
bit 12: C1 PROC CTRL SENSOR (TC7)
bit 13: C2 PROC CTRL SENSOR (TC8)
bit 14: Not Used
bit 15: BUFFER LINE TC FAULT (TC16)
bit 16: OPTION1 TC FAULT (TC17)
bit 17: OPTION2 TC FAULT (TC18)
bit 18: OPTION3 TC FAULT (TC19)
bit 19: BP CHK CTRL (TC6)
bit 20: CLD JUNCTION SENSOR (TC21)
bit 21: Not Used
bit 22: Not Used
bit 23: OPTION4 TC FAULT (TC20)
bit 24: Not Used
bit 25: Not Used
bit 26: DSCHRG PRESS SENSOR (PT2)
bit 27: SUCT PRESS SENSOR (PT1)
bit 28: Not Used
bit 29: Not Used
bit 30: Not Used
bit 31: Not Used
Default Value: N/A
Security Level Get: 1 Set: N/A
Example: “FAULT_STATUS,?”
Interface RS-232, Ethernet
Purpose: Get the current alarm or warning status.
Parameter(s): [n], [nnnnnnnn],?
[n] is
0 = Equipment operation alarms
1 = Sensor range alarms
2 = Performance Monitor alarms
[nnnnnnnn] is a 32 bit number / bitfield that represents the state of the
alarm. Displays the hex number that represents the state of the 6 bits of
one of the bitfields returned by ACTIVE_ALARMS command