Host Commands Polycold Cryochiller
System Configuration Commands Installation and Operation Manual
Brooks Automation
7-12
214072 Revision B
PROFIBUS_PARAMS Profibus parameters
HMI_TIMEOUT HMI Timeout
Purpose: Bet the profibus speed and Get/Set Profibus
Parameter(s): SPEED,? or ,ADDRESS,? or ,ADDRESS,[address]
[speed] is READ ONLY
-0 = 9.6k
-1 = 19.2k
- 2 = 45.45k
- 3 = 93.75k
- 4 = 187.5k
- 5 = 500k
-6 = 1.5M
-7 = 3M
-8 = 6M
-9 = 12M
[address] is 1 - 125.
Return Value: [A] or [A][speed] or [A][address]
Default Value: [speed] = 0 [address] = 7
Min: [speed] = 0 Read Only
[address] = 1
Max: [speed] = 9 Read Only
[address] = 125
Security Level Get: 1 Set: 3
Example: “PROFIBUS_PARAMS,SPEED,?” Get the Profibus speed
Interface RS-232, Ethernet
Purpose: Get/Set the timeout of the HMI menu to previous menu
Parameter(s): ? or [nnnn]
[nnnn] is HMI timeout in seconds
Return Value: [A] or [A][nnnn]
Default Value: 300
Min: 30 Max: 1800
Security Level: Get: 1 Set: 3
Example: “HMI_TIMEOUT,?” get the current HMI timeout setting
Interface RS-232, Ethernet