Host Commands Polycold Cryochiller
System Operational State Commands Installation and Operation Manual
Brooks Automation
7-16
214072 Revision B
System Operational State Commands
SYSTEM_CONTROL_MODE_DEF Default System Control Mode
SYSTEM_CONTROL_MODE System Control Mode
Purpose: Get/set the default system control mode on startup. This command does
not actually change the mode.
Parameter(s): ? or [mode]
[mode] is
0 = local mode
1 = remote Ethernet
2 = remote RS232
3 = remote RS485 (Not supported)
4 = remote 24V DI/DO Remote
5 = remote DeviceNet
6 = remote Profibus
Return Value: [A] or [A][n]
Default Value: Base Units: 2 Units with Ethernet: 1
Security Level Get: 1 Set: 3
Example: “SYSTEM_CONTROL_MODE_DEF,0” Set the system default control
mode to local mode.
Interface RS-232, Ethernet
Purpose: Get/set the system control mode.
Parameter(s): ? or [mode]
[mode] is
0 = local mode
1 = remote Ethernet
2 = remote RS232
3 = remote RS485
4 = remote 24V DI/DO Remote
5 = remote DeviceNet
6 = remote Profibus
8 = Default Control Mode (SYSTEM_CONTROL_MODE_DEF)
Return Value: [A] or [A][n]
Default Value: Determined by command SYSTEM_CONTROL_MODE_DEF
Security Level Get: 1 Set: 2
Example: “SYSTEM_CONTROL_MODE,0” Set the system to local mode.
Interface RS-232, Ethernet