Host Commands Polycold Cryochiller
System Sensor Readings Installation and Operation Manual
Brooks Automation
7-26
214072 Revision B
BUFFER_LINE_TEMP Buffer Line Temperature
WARM_CTL_TEMP Rapid Balance Pressure Check Control Temperature
Units: Degrees Celsius (signed)
Security Level Get: 1 Set: N/A
Example: “COLD_JUNCT_TEMP,?”
Interface RS-232, Ethernet
Purpose: Get the buffer line temperature. (TC16)
Parameter(s): ?
Return Value: [A][nnn.n]
Default Value: N/A
Units: Degrees Celsius (signed)
Security Level Get: 1 Set: N/A
Example: “BUFFER_LINE_TEMP,?”
Interface RS-232, Ethernet
Purpose: Get the Rapid Balance Pressure Check control temperature sensor
value. (TC6)
Parameter(s): ?
Return Value: [A][nnn.n]
Default Value: N/A
Units: Degrees Celsius (signed)
Security Level Get: 1 Set: N/A
Example: “WARM_CTL_TEMP,?”
Interface RS-232, Ethernet