EasyManua.ls Logo

BRUEL & KJAER PULSE 3560-C - 12-Ch. Dyn-X Input Modules Types 3040, 3040-B

BRUEL & KJAER PULSE 3560-C
162 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PULSE Multi-analyzer System Type 3560-B/C/D/E Installation and IDAe Hardware82
6.11 12-ch. Dyn-X Input Modules Types 3040, 3040-B
This module comes in two versions: Type 3040
with BNT/BNC connectors on the input channels
and Type 3040-B with two 37-pin D-sub connec-
tors. For input of signals in one range to 10 V
peak
(dynamic range 160 dB) over the frequency range
to 25.6 kHz.
6.11.1 Common Features
Input may be grounded or floating, controlled by
PULSE software.
LEDs
Measuring: Indicates a measurement in progress
or the downloading of software to the module.
Input 1 to 12: For each channel, an LED indicates
the status of the channel (activated: green; over-
load: red). See section 7.8. On Type 3040, the
LED surrounds the connector.
6.11.2 Type 3040
Inputs 1 and 7: BNT connector (see Fig.7.3 on page 89) for direct connection of a voltage
signal or for connection of CCLD transducers. The connectors also provide power for tacho
probes.
Inputs 2 to 6 and 8 to 12: BNC connectors (see Fig.7.4 on page 89) for direct connection
of a voltage signal or for connection of a CCLD transducer, for example, DeltaTron
®
accel-
erometers.
6.11.3 Type 3040-B
Two 37-pin D-sub connectors for inputs 1 to 6, and 7 to 12. See Fig.7.8 on page 92.
The module can provide polarization voltage to all channels simultaneously and support
IEEE 1451.4 capable transducers with TEDS. It provides a supply for CCLD transducers,
for example, DeltaTron
®
accelerometers, and for microphone preamplifiers.

Table of Contents

Related product manuals