BVT3000 Version 003 BRUKER 13 (67)
Front panel connectors
Front panel connectors 1.7
Heater connector 1.7.1
The heater is plugged into this connector. A safety thermocouple is located close
to the heater resistance in order to detect an overheating in the case of a missing
gas flow for example.
Figure 1.5. Heater connector (Front view)
Table 1.2. Heater connector pin assignment
PIN 7 pins NMR 8 pins ESR
1 heater + heater +
2 heater + NC
3 security thermocouple + heater -
4 security thermocouple - security thermocouple +
5 heater - security thermocouple -
6 heater - GND
7 gnd heater +
8 / heater -
43
2
16
57
5
4
1
7
8
3
6
2
NMR ESR