BVT3000 Version 003 BRUKER 29 (67)
List of commands
List of commands 4.5
Table 4.2. List of commands
COMMANDS R W COMMENT
AF X X reads / writes gas flow delivery
CM X starts a ram test of the microcontroller (for test only)
CO X X reads / writes communication speed (Interface <-> Eurotherm)
DL X X reads the download transfer status/ initialises a download transfer
DT X DAC check (for test only)
ES X reads the error status
HP X X reads / writes heater power state ('1' or '0')
IS X reads interface status
NH X X reads / writes LN2 heater power level
NP X X reads / writes LN2 heater power state ('1' or '0')
P1 X X reads / writes port 1 (for test only)
P2 X X reads / writes port 2 (for test only)
P3 X reads port 3 (for test only)
P4 X reads port 4 (for test only)
SV X reads interface version (software, hardware and installed options)
RB X reads BBIS memory content
WB X writes to a BBIS memory location
WR X writes a record to the BVT3000
XR X transmit a hexadecimal record to the host