BVT3000 Version 003 BRUKER 25 (67)
3
Configuration 3
Sensor selection 3.1
0
The BVT3000 can be used with different types of sensors:
Thermocouple T or E for NMR, and K for ESR spectrometer
BTO2000
PT100 sensor
Warning: Never connect two types of sensors at a same time on the
BVT3000.
Eurotherm 902 configuration 3.2
0
The EUROTHERM 902 controller must be configured to work with the right type of
sensor.
The configuration is done with the Eurotherm 902 controller keypad (or by soft-
ware in the EDTE program on NMR spectrometer).
Proceed as follows:
1. Switch off the main power.
2. Press the two left most keys : at same time while turn-
ing the power on.
3. The messages CONF and UCONF appear on the display.
4. Press the scroll button until ICONF is displayed.
5. Press the left button until C1 appears.
6. Now press the increment button or decrement button :
the 4 digit code is displayed. This code must be changed to select the new
sensor.