BVT3000 Version 003 BRUKER 37 (67)
Authorised functions
ES (Error status) 4.8.1
Read
Syntax: EOT0000ES ENQ
Response
:STXES val ETX BCC
Description
: Allows the user to get information about the last six errors.
Explanations
: At each «ES» request, the last happened error code is sent and
then reset. If more than six errors are memorised, the oldest error code is repla-
ced by the new one. To erase all errors, the user must sent «ES» requests until
response is «ES0».
The returned value «Val» informs the user about errors. The different error codes
are given below.
Write
Syntax: EOT0000STXES val ETX BCC
Response
:AlwaysNACK.
Table 4.4. Error status description
VAL SIGNIFICATION COMMENT
0 NOERROR no error in command
1 SYNTAX unknown command/syntax error
2 checksum checksum error
3 erasefail flash eprom erase error
4 programmfail flash eprom program error
5 wrongrecordtype no intel hex record
6 wrongaddress program address out of range
7 wrongchecksum checksum error intel hex
8 wrongtransmissioncheck wrong eof record
9 wrongdatacount byte counter error
10 noappsw no application software
11 nobbis no BBIS available
12 bbiscs1 BBIS checksum error block 1
13 bbiscs2 BBIS checksum error block 2
14 bbiscs3 BBIS checksum error block 3
15 bbiscs4 BBIS checksum error block 4