BVT3000 Version 003 BRUKER 41 (67)
Authorised functions
NP (Nitrogen heater power) 4.8.5
Write
Syntax: EOT0000STXNP state ETX BCC
Response
:ACK
Description
: This command allows the nitrogen heater power to be switched ON
or OFF.
Rules
: State can be «0» or «1».
«1» switch liquid nitrogen heater ON.
«0» switch liquid nitrogen heater OFF.
WARNING
: Space characters are not allowed.
Read
Syntax: EOT0000NP ENQ
Response
:STXNP State ETX BCC
Description
: This command allows the nitrogen power to be read back.
Rules
: State can be «0» or «1».
A equal «1» means that liquid nitrogen heater is ON
A equal «0» means that liquid nitrogen heater is OFF
NB: After power on the nitrogen heater power is at «0»