46 (67) BRUKER BVT3000 Version 003
Remote interface control
RB (Read BBIS) 4.8.10
Write
Syntax: EOT0000STXRB adr_e2prom A1 A0 ETX BCC
Response
:STXRB>D0D1ETX
NACK if command can't issue.
Description
: RB command allows a single byte to be read in a BBIS E2PROM
Rules
: A1and A0 are values from '0' up to 'F' representing the byte address
in the E2PROM.
adr_e2prom is a value from '0' up to '7' representing the I2C bus address of
the E2PROM
- 0: Address unused
- 1: BVT3X00 motherboard address
- 2: BVTB3500 (Booster) address
- 3: Address unused
- 4: Address unused
- 5: Address unused
- 6: Address unused
- 7: Address unused
All other values generates a NACK response